11-4 Appendix 2
4/4
QIS 12J05D02-01E
3.7 Atmospheric Pressure Indication Check
Following Section 3.6, press the [ENT] key until the message display shows “PRESS.”
Check that the data display indication at this time is within the range from 900 (hPa) to
1100 (hPa).
Press the [ENT] key. The message display will show “READY.”
Continue to press the [ENT] key to return to normal measurement mode.