IM 12J05D02-01E

11-4 Appendix 2

4/4

QIS 12J05D02-01E

3.7 Atmospheric Pressure Indication Check

Following Section 3.6, press the [ENT] key until the message display shows “PRESS.”

Check that the data display indication at this time is within the range from 900 (hPa) to

1100 (hPa).

Press the [ENT] key. The message display will show “READY.”

Continue to press the [ENT] key to return to normal measurement mode.