IM 12J05D02-01E

2-2 Specification

J. Contact outputs

- General : Four (4) SPDT relay contacts

with LED indicators. For S1,

S2, and S3, the LED is on

when relay is powered.

NOTE:

For S4 (FAIL) LED lights when

power is removed (Fail safe).

Contact outputs configurable

for hysteresis and delay time.

- Capacity

: Maximum values 100 VA,

250 V AC, 5 Amps.

Maximum values 50 Watts,

250 V DC, 5 Amps.

- Status

: High/Low process alarms,

selected from process

parameters and temperature.

Contact output is also available

to signal “Hold Active

Contact S1 to S3

Contact S4

Status

LED

NO

NC

LED

ON

NC

Alarm,
FAIL,
OFF

On

Closed

Open

On

Open

Closed

Alarm,
FAIL,
OFF

Off

Open

Closed

Off

Closed

Open

Power
OFF

Off

Open

Closed

Off

Open

Closed

- Control function

: On/Off

PI pulsed Proportional duty cycle control

with integral term.

PI frequency Proportional frequency control

with integral term. In addition

wash cleaning control signal on

S3, and FAIL alarm for system

and diagnostic errors on S4

K. Contact input: Wash start or input remote range

change (either choice)

On resistance: 10Ω or less

Off resistance: 100kΩ or more

On time:

0.5 sec. or more

L. Power supply

Supply voltage rating

: 100, 115, 230 VAC

Applicable range

: 85 to 115, 97.8 to 132.2,

195.5 to 264.5 VAC

Supply frequency rating

: 50 / 60 Hz

Applicable range

: 50 Hz ± 5% / 60 Hz ± 5%

Power consumption

: Maximum 10 VA for steady

operation

M. Safety and EMC conforming standards

,

Safety:

conforms to EN 61010-1

EMC:

EN 61326-1 Class A, Table 2 (For

use in industrial locations) (Note 1)

EN 61326-2-3

EN 61000-3-2 Class A

EN 61000-3-3

EMC Regulatory Arrangement in

Australia and New Zealand

EN 55011 Class A, Group 1

Korea Electromagnetic Conformity

Standard Class A

한국 전자파적합성 기준

Installation altitude: 2000 m or less

Category based on IEC 61010: II (Note 2)

Pollution degree based on IEC 61010:2 (Note 2)

Note 1: This instrument is a Class A product,

and it is designed for use in the

industrial environment. Please use

this instrument in the industrial

environment only.

A급 기기 (업무용 방송통신기자재)
 이 기기는 업무용(A급) 전자파적합기기로서 판매자 또는
사용자는 이 점을 주의하시기 바라며, 가정외의 지역에서
사용하는 것을 목적으로 합니다.

Note 2: Installation category, called over-

voltage category, specifies impulse

withstand voltage. Category II is for

electrical equipment.

Pollution degree indicates the degree

of existence of solid, liquid, gas or

other inclusions which may reduce

dielectric strength. Degree 2 is the

normal indoor environment.

N. Shipping details

: Package size w x h x d

290 x 300 x 290 mm.

11.5 x 11.8 x 11.5 in.

Packed weight approx.

2.5 kg (5lb).

2-2. Operating specifications

A. Performance : DO (at t process = 25 °C)

- Linearity : ± 0.03 mg/l or ± 0.5%FS,

whichever is greater

- Repeatability: ± 0.03 mg/l or ± 0.5%FS,

whichever is greater

- Accuracy : ± 0.05 mg/l or ± 0.5%FS,

whichever is greater