3. wiring the power supply, 3-1. general precautions, 3. wiring the power supply -5 – Yokogawa DO402 Dissolved Oxygen Analyzer User Manual

Page 25: 3-1. general precautions -5

Advertising
background image

IM 12J05D02-01E

Installation and wiring 3-5

Figure 3-6. System configuration

3-3. Wiring the power supply

3-3-1. General precautions

Make sure the power supply is switched off. Also, make sure that the power supply is correct for the

specifications of the EXA and that the supply agrees with the voltage specified on the textplate. Remove

the front cover by unscrewing the four screws to check this textplate on the top of the display board.

Local health and safety regulations may require an external circuit breaker to be installed. The instru-

ment is protected internally by a fuse. The fuse rating is dependent on the supply to the instrument. The

250 VAC fuses should be of the “time-lag” type, conforming to IEC60127.

The internal fuse is located next to the power terminals (in the lower right hand corner).

Use only a fuse of the specified current, voltage and type ratings to prevent fire. For fuse replacement,

refer to Section 7-3, “Fuse Replacement.”

DANGER

FRONT GLANDS

Sensor

Output

signals

RS485

Power

Contact

output

Contact

output

REAR GLANDS

Contact intput

S1

S2

S3

S4/FAIL