Yokogawa DO402 Dissolved Oxygen Analyzer User Manual
Page 53

IM 12J05D02-01E
Parameter setting 5-17
Code
Display
Function
Function detail
X
Y
Z
Default values
Temperature Measuring Functions
10
*T.SENS
Temp. Comp.
Pt100 RTD
0
Pt1000 RTD (DO30)
1
1
PB36 = 2k2 NTC
2
22k NTC (SM31) (DO70G)
3
11
*T.UNIT
Temp units
Celsius
0
0
Fahrenheit
1
12
*T.ADJ
Temp adjust
Adjustment +/- 7.5 °C or +/- 13.5 °F
13
*T.MAN
Manual TC
Automatic temperature comp
0
0
Manual temp compensation
1
25 °C/ 77 °F
14-19
Not used
Code
Display
Function
Function detail
X
Y
Z
Default values
Calibration Settings
20
*∆T.SEC
*∆mg/l
Stabilization
Step
Stabilisation time range 10- 600 s
60 s
DO step change range 0-50 mg/l
or 0-300 %
0.05 mg/l
21
*0.CAL
Zero cal.
Zero calibration disabled
0
0
Zero Calibration enabled
1
22
*ZERO
Zero entry
Galvanic:
In case that zero check is activated in code 02
within *I.CELL value set in code 02
In case that zero check is not activated in code
02 ±19.99 µA
00.00 µA
000.0 nA
Polarographic:
In case that zero check is activated in code 02
within *I.CELL value set in code 02
In case that zero check is not activated in code
02 ±199.9 nA
*SLOPE
Slope entry
Minimum setting
(*I.CELL set in code 01)
× (*SL.MIN set in code 02)
Maximum setting
(*I.CELL set in code 01)
× (*SL.MAX set in code 02)
3.75 µA (50 nA)
23-29
Not used