IM 12J05D02-01E

Parameter setting 5-17

Code

Display

Function

Function detail

X

Y

Z

Default values

Temperature Measuring Functions

10

*T.SENS

Temp. Comp.

Pt100 RTD

0

Pt1000 RTD (DO30)

1

1

PB36 = 2k2 NTC

2

22k NTC (SM31) (DO70G)

3

11

*T.UNIT

Temp units

Celsius

0

0

Fahrenheit

1

12

*T.ADJ

Temp adjust

Adjustment +/- 7.5 °C or +/- 13.5 °F

13

*T.MAN

Manual TC

Automatic temperature comp

0

0

Manual temp compensation

1

25 °C/ 77 °F

14-19

Not used

Code

Display

Function

Function detail

X

Y

Z

Default values

Calibration Settings

20

*∆T.SEC

*∆mg/l

Stabilization

Step

Stabilisation time range 10- 600 s

60 s

DO step change range 0-50 mg/l

or 0-300 %

0.05 mg/l

21

*0.CAL

Zero cal.

Zero calibration disabled

0

0

Zero Calibration enabled

1

22

*ZERO

Zero entry

Galvanic:

In case that zero check is activated in code 02

within *I.CELL value set in code 02

In case that zero check is not activated in code

02 ±19.99 µA

00.00 µA

000.0 nA

Polarographic:

In case that zero check is activated in code 02

within *I.CELL value set in code 02

In case that zero check is not activated in code

02 ±199.9 nA

*SLOPE

Slope entry

Minimum setting

(*I.CELL set in code 01)

× (*SL.MIN set in code 02)

Maximum setting

(*I.CELL set in code 01)

× (*SL.MAX set in code 02)

3.75 µA (50 nA)

23-29

Not used