2-18

• Single-gauge three-wire method

R

R

R

Rg

r

r

r

E

e

R = fixed resistor
r = resistance of leadwire
Rg = resistance of strain gauge
e = output voltage developed across bridge
E = voltage imposed across bridge

Jumper setup switch

DU500-12/DU500-13

DU500-14

ON

OFF

No.1
No.2

No.3

No.4
No.5

A(+)

B(L)

C(-V)

D(H)

No.1

ON

No.2

ON

No.3

OFF

No.4

ON

No.5

OFF

Rg

1

2

3

4

5

6

7

8

• Adjacent-side two-gauge method

Jumper setup switch

DU500-12/DU500-13

DU500-14

ON

OFF

No.1
No.2

No.3

No.4
No.5

A(+)

B(L)

C(-V)

D(H)

No.1

ON

No.2

ON

No.3

OFF

No.4

OFF

No.5

ON

Rg1

Rg2

R

R

E

e

Rg1

Rg2

Rg1

Rg2

R = fixed resistor
r = resistance of leadwire
Rg = resistance of strain gauge
e = output voltage developed across bridge
E = voltage imposed across bridge

1

2

3

4

5

6

7

8

Rg2

Rg1

• Opposed-side two-gauge method

Jumper setup switch

DU500-12/DU500-13

DU500-14

ON

OFF

No.1
No.2

No.3

No.4
No.5

A(+)

B(L)

C(-V)

D(H)

No.1

ON

No.2

OFF

No.3

ON

No.4

OFF

No.5

ON

Rg1

Rg2

R

R

Rg2

Rg1

E

e

Rg1

Rg2

R = fixed resistor
r = resistance of leadwire
Rg = resistance of strain gauge
e = output voltage developed across bridge
E = voltage imposed across bridge

The 319300 bridge box does
not support the opposed-
side two gauge method.

2.6 Connecting the Signal Lines