Warning – Yokogawa DA100 User Manual

Page 46

Advertising
background image

IM DA100-01E

2-25

2

Installation and Wiring

2.8 Connecting the Power Cord and Turning the

Power ON/OFF

When Using the Accessory Power Cord

Before connecting the power supply to the DA100/DS400/DS600, make sure to comply with the
following warnings. Failure to do so may cause electric shock or damage to the instrument.

WARNING

• Connect the power cord only after confirming that the voltage of the

power supply matches the rated electric power voltage for this
instrument.

• Connect the power cord only after confirming that the power switch of

DA100/DS400/DS600 is turned OFF.

• Only use power cords which are supplied with the DA100/DS400/

DS600 by YOKOGAWA in order to prevent electric shock or fire
hazard.

• Always use protective ground to prevent electric shock. Connect the

accessory power cord of the DA100/DS400/DS600 to a power outlet
with protective grounding. Do not use the function grounding terminal
(below the power switch) as protective grounding.

• Never use an extension cord that does not have protective grounding,

otherwise the protection feature will be negated.

Connecting Procedure

1 Verify that the power switch of the DA100/DS400/DS600 has been turned OFF.
2 Connect the plug of the accessory power cord to the power connector of the DA100/DS400/DS600.
3 Plug the other end of the power cord into an AC outlet that meets the following specifications.

The AC outlet must be of a 3-pin type with a protective grounding terminal. Make sure to
apply the clamp filter (standard accessory) to the power cord as shown below. This is to
eliminate the electric emmission.

Rated supply voltage

:100 to 240 VAC

Permitted supply voltage

:90 to 250 VAC

Rated supply voltage frequency

:50/60 Hz

Maximum power consumption:

DA100 Stand-alone type (with 6 modules connected):

in case of 100VAC input: max. 55VA
in case of 240VAC input: max. 70VA

DA100 Expandable type (with 4 modules connected):

in case of 100VAC input: max. 45VA
in case of 240VAC input: max. 55VA

DS400 Subunit (with 4 modules connected):

in case of 100VAC input: max. 45VA
in case of 240VAC input: max. 55VA

DS600 Subunit (with 6 modules connected):

in case of 100VAC input: max. 55VA
in case of 240VAC input: max. 70VA

Power switch

Power connector

Status indicator

Power Cord

Clamp filter (Standard Accesory: A1179MN)