3 trouble-shooting, Trouble-shooting -8 – Yokogawa DA100 User Manual

Page 67

Advertising
background image

IM DA100-01E

3-8

3.3

Trouble-Shooting

• When a message appears on the screen, first refer to the list of error messages described on page

3-4.

• When servicing is necessary, contact your nearest sales representative. Addresses may be found

on the back cover of this manual.

Description

Cause

Action

Reference
Page

No power-up; the status

The used power supply lies

Use the correct power supply.

2-25

indicator does not light.

outside the permissible range.

The status indicator is

The communication module

Configure the communication

2-4

flashing in 1-second

is not configured properly.

module properly.

intervals.

The status indicator is

An internal error occurred.

Turn the power OFF and

2-25

flashing in an interval

ON again.

other than 1-second.

Connected sub-units are

The power of the corresponding

Turn the power ON.

2-25

not recognized.

sub-unit has been turned OFF.
The extension cables of the

Connect the extension cables

2-13

corresponding sub-unit are not

properly.

connected properly.

Connected modules are

The corresponding module is not

Connect the module properly.

2-4

not recognized.

properly connected.

Measurement values are

The input does not conform

Verify the input

Chapter 4

not correct.

specifications.

specifications.

The measurement mode is not

Select a suitable measurement

*1

suited for this input.

mode.

The measurement range/span/scale Select a suitable measurement

*1

setting is not suited for this input.

range/span/scale.

Noise interference.

Apply anti-noise measures.

2-29

Input wiring is not correct.

Wire the input properly.

2-15

RJC settings are not correct.

Enter the correct RJC settings. *1

The environmental temperature

Make sure the environmental

2-2

is outside the specifications.

temperature matches the
specifications.

*1 Refer to the User’s Manual for DAQ Software 32.