IM DA100-01E

4-18

4.5

Specifications of Strain Input Module

Style Number: S5
Model Code, Number of Input Channels, Terminal Configuration, Shortest
Measurement Interval and Values of Built-in Resistors

Model Code

Number of Input

Terminal

Shortest Measurement

Built-in

Channels

Configuration

Interval

Resistor

DU500-12

10*

Clamp

0.5 s

120

DU500-13

10*

Clamp

0.5 s

350

DU500-14

10*

NDIS**

0.5 s

External

*: Requires the space of two slots.
**: Recommended by the Japanese Nondestructive Inspection Association

Method of Input

Non-balanced floating input, with isolation between channels (separated channels)

Resolution of A/D Conversion

Equivalent to reading on 20000 full scale

Integral Time of A/D Conversion

Manual or automatic selection between 20 ms (50 Hz), 16.7 ms (60 Hz) and 100 ms (10 Hz)
However, AUTO does not function if the instrument is the stand-alone DC power supply model,
or the subunit (DS400/DS600) of the DC power supply model. (Selecting “AUTO” will set the
A/D integration time to 20 ms (50 Hz)).

Types of Input

Signals from strain gauge or strain-gauge sensor

Connection Methods

Single-gauge, opposed-side two-gauge, adjacent-side two-gauge or four-gauge configuration
(Configured with built-in DIP switches for DU500-12 and DU500-13)

Applicable Gauge Resistance

100 to 1,000

The DU500-12 and DU500-13 modules contain 120-

Ω and 350- Ω resistors, respectively.

Bridge Voltage

Fixed to 2 V DC (

±5% accuracy with a correction function)

Gauges

2.00 (with scaling capability)

Balancing

Automatic

Range of Balancing

Equivalent to

±10,000 µε (for single-gauge method)

Measuring Ranges and Accuracies
Reference operation conditions

temperature: 23

±2°C; humidity: 55 ±10% RH; supply voltage 90 to 250 VAC (AC power supply)

/10 to 32 VDC (DC power supply); supply frequency: 50/60

±1% (AC power supply); warm-up

time: at least 30 minutes; when operating, the system must not adversely affect the operation of
other equipment.

Gauge Method

Measuring Range

Accuracy

Resolution

Single-gauge

–2000 to 2000

µε

0.5% of Range

0.1

µε

–20000 to 20000

µε

0.3% of Range

1

µε

–200000 to 200000

µε

0.3% of Range

10

µε

Two-gauge

–1000 to 1000

µε

0.5% of Range

0.1

µε

10000 to 10000

µε

0.3% of Range

1

µε

–100000 to 100000

µε

0.3% of Range

10

µε

Four-gauge

–500 to 500

µε

0.5% of Range

0.1

µε

–5000 to 5000

µε

0.3% of Range

1

µε

–50000 to 50000

µε

0.3% of Range

10

µε

The Accuracies and Resolutions of the DU500-12 and DU500-13 models, do not include the influence of the
internal-and contact-resistance of the jumper setup switch.