4-10

Measurement range and accuracy

Note that RTD input is not possible for the DCV/TC/DI input module.

Reference operation conditions

temperature: 23

±2°C; humidity: 55 ±10% RH; supply voltage 90 to 250 VAC (AC power supply)

/10 to 32 VDC (DC power supply); supply frequency: 50/60

±1% (AC power supply); warm-up

time: at least 30 minutes; when operating, the system must not adversely affect the operation of
other equipment.

Input

Measuring(digital display)

DC Voltage

TC
(Note that Accuracy
of reference junction
compensation is not
considered).

RTD

High resolution RTD

Contact

20mV
60mV
200mV
2V
6V
20V
50V
R

S
B
K

E
J
T
L
U
N
W
KPvsAu7Fe
Pt100(1mA)
Pt100(2mA)
JPt100(1mA)
JPt100(2mA)
Pt50(2mA)
Ni100(1mA)
SAMA
Ni100(1mA)DIN
Ni120(1mA)
J263*B
Cu10 GE
Cu10 L&N
Cu10 WEED
Cu10 BAILEY
Pt100(1mA)
Pt100(2mA)
JPt100(1mA)
JPt100(2mA)
Voltage input

Contact input

-20.000 to 20.000mV
-60.00 to 60.00mV
-200.00 to 200.00mV
-2.0000 to 2.0000V
-6.000 to 6.000V
-20.000 to 20.000V
-50.00 to 50.00V
0.0 to 1760.0

°C

0.0 to 1760.0

°C

0.0 to 1820.0

°C

-200.0 to 1370.0

°C

-200.0 to 800.0

°C

-200.0 to 1100.0

°C

-200.0 to 400.0

°C

-200.0 to 900.0

°C

-200.0 to 400.0

°C

0.0 to 1300.0

°C

0.0 to 2315.0

°C

0.0 to 300.0K
-200.0 to 600.0

°C

-200.0 to 250.0

°C

-200.0 to 550.0

°C

-200.0 to 250.0

°C

-200.0 to 550.0

°C

-200.0 to 250.0

°C

-60.0 to 180.0

°C

-70.0 to 200.0

°C

0.0 to 300.0K
-200.0 to 300.0

°C

-140.00 to 150.00

°C

-70.00 to 70.00

°C

-140.00 to 150.00

°C

-70.00 to 70.00

°C

Off for a voltage of less
than 2.4V.
On for a voltage of 2.4V
or more.(TTL)
On/Off of contact

Type

Measurement accuracy

Maximum

resolution

±(0.05% of rdg + 5digits)

±(0.05% of rdg + 2digits)

±(0.05% of rdg + 2digits)

±(0.05% of rdg + 2digits)

±(0.05% of rdg + 2digits)

±(0.05% of rdg + 2digits)

±(0.05% of rdg + 2digits)

±(0.05% of rdg + 1°C)
However,R,S:0 to 100

°C,±3.7°C

100 to 300

°C,±1.5°C

B:400 to 600

°C,±2°C accuracy

less than 400

°C is not specified.

±(0.05% of rdg +0.7°C)
However,K attains an accuracy of
±(0.05% of rdg. +1°C) within the
range between -200 to -100

°C.

±(0.05% of rdg +0.5°C)
However,J and L attain an accuracy of
±(0.05% of rdg. +0.7°C) within the
range between -200 to -100

°C.

±(0.05% of rdg. +0.7°C)

±(0.05% of rdg. +1°C)

±(0.05% of rdg. +0.7K)

±(0.05% of rdg. +0.3°C)

±(0.05% of rdg. +0.3°C)

±(0.05% of rdg. +0.3°C)

±(0.05% of rdg. +0.3°C)

±(0.05% of rdg. +0.3K)

±(0.2% of rdg. +0.7°C)

±(0.05% of rdg. +0.3°C)

±(0.05% of rdg. +0.3°C)

±(0.05% of rdg. +0.3°C)

±(0.05% of rdg. +0.3°C)
Off for a voltage of less than 2.4V.

On for a voltage of 2.4V or more.
(TTL)
On/Off of contact

Measurement range

1

µV

10

µV

10

µV

100

µV

1mV
1mV
10mV

0.1

°C

0.1K

0.1

°C

0.1K

0.1

°C

0.01

°C

*

1

*

1

*

1

*

1

*

1

*

1

*

1

*

2

*

2

*

3

*

4

*

5

*

5

*

5

*

5

*

5

*

6

*

6

*

7

*

8

*

8

*

8

*

8

*

5

*

5

*

5

*

5

*1 : R,S,B,K,E,J,T:ANSI,IEC 584,DIN IEC 584,JIS C 1602-1981
*2 : L:Fe-CuNi,DIN43710,U:Cu-CuNi,DIN 43710
*3 : N:Nicrosil-Nisil,IEC 584,DIN IEC 584
*4 : W:W·5%RE-W·26%Re (Hoskins Mfg Co)
*5 : Pt50:JIS C 1604-1981,JIS C 1606-1986
Pt100:JIS C 1604-1989,JIS C 1606-1989,IEC 751,DIN IEC 751
JPt100:JIS C 1604-1981,JIS C 1606-1989

*6 : SAMA/DIN
*7 : McGRAW EDISON COMPANY
*8 : ranges to which accuracy applies:
Cu10 GE:-84.4 to 170.0

°C,

Cu10 L&N:-75.0 to 150.0

°C,

Cu10 WEED:-20.0 to 250.0

°C,

Cu10 BAILEY:-20.0 to 250.0

°C

Noise rejection

rejection by integration type A/D, lowpass filter, or moving averaging.

Cut-off frequency

50/60/10Hz

Input resistance

Min. 10M

Ω at 2V DC or lower, thermocouple range

Approx. 1M

Ω at 6V DC or higher.

4.2 Universal Input Module and DCV/TC/DI Input Module

Advertising