Yokogawa DA100 User Manual

Page 96

Advertising
background image

IM DA100-01E

4-27

4

Specifications

Transfer distance

max. 15 m

Connector

D-sub 25pin

Handshaking

hardware

: transmission and reception control by ‘DTR’’ ‘RTS’, ‘CTS’ signal enabled.

Software

: transmission control by ‘XON’ and ‘XOFF’ enabled.

Capacity of receiving buffer

200 bytes

Escape sequence

for reception only

Functions

· Talker functions

Output of measurement data (ASCII, binary) and setting parameters (ASCII)

· Listener functions

Setting of measurement conditions, control of measurement start and stop (excluding the setting
and control of power on/off), specifying causes of ‘ESC S’ (output of a status byte)

Contents of ‘status’

Syntax error, completion of A/D conversion and data acquisiton error (occurs only when using
optional math function)

External dimensions

Approx. 57 (W)

× 137 (H) × 53.8 (D) mm

Weight

0.25kg

RS-422-A/RS-485 Module

Style Number: S8
Electrical and Mechanical Specifications

Compatible with EIA RS-422-A and EIA RS-485

Connection

1 : n multi-drop (n = 16 for RS-422-A; 31 for RS-485)

Communication

Half-duplex

Synchronization

Start-stop (by means of start and stop bits)

Baud Rates

Selected from 300, 600, 1200, 2400, 4800, 9600, 19200, 38400 bps

Start Bit

Fixed to one bit

Data Length

Selected from 7 and 8 bits

Parity

Selected from Even, Odd and None

Stop Bits

Selected from 1 and 2 bits

Connector

Six-line, screw-terminal (with 4-mm screws)

Minimum Response Time

Selected from 0, 10, 20, 50 and 100 ms

Length of Receive Buffer

250 bytes

Escape Sequence

Trigger, status call, open, close

Electrical Characteristics

SDA, SDB, RDA, RDB and SG lines, where the signal-line terminal is functionally isolated from
the module's internal circuitry.

4.10 Communication Interface Module

Advertising