7 specifications of digital input module, Specifications of digital input module -22 – Yokogawa DA100 User Manual

Page 91

Advertising
background image

IM DA100-01E

4-22

4.7 Specifications of Digital Input Module

Style Number: S8
Model, Number of Channels, Terminal Type and Minimum Measurement Interval

Model

Number of Channels

Terminal Type

Minimum Measurement Interval

DU700-11

10

Screw

05s

Input method

Floating unbalanced input, each channel mutually isolated (channel independent)

A/D resolution

±20000

A/D integration time

20 ms (50 Hz), 16.7 ms (60 Hz), 100 ms (10 Hz) or auto
(Except, auto switch does not function on the stand-alone DC power supply model or the subunit
(DS400/DS600) of a DC power supply model.)

Minimum measurement interval

Low-pass filter OFF

Cut-off frequency

50/60Hz

10Hz

Minimum measurement interval

05 s

4 s

Low-pass filter ON

Cut-off frequency

50/60Hz

10Hz

Minimum measurement interval

3 s

12 s

Normal operating temperature/humidity

20 to 80% RH for –10 to 40

°C, 10 to 50% RH for 40 to 50°C, 5 to 30% RH for 50 to 60°C (no

condensation)

Magnetic field

400A/m max (50/60Hz)

Maximum allowable input voltage

CONT (Contact input) :

±10 V DC

LEVEL (Voltage input) :

±60 V DC

Common mode noise voltage

250 V AC rms (50/60 Hz)

Maximum noise between channels

150 V AC rms (50/60 Hz)

Action

At normal operating temperature/humidity
Voltage input (LEVEL) :

Off for a voltage of less than 24V.
ON for a voltage 2.4 V or more.

Contact input (CONT) :

On/Off of contact.*

*: Contact resistance

ON ; 2 k

Ω or less, OFF ; 100 k Ω or more

Noise rejection

Rejection by integration type A/D, lowpass filter

Cut-off frequency

50/60/10Hz

Input resistance

CONT : Min. 10 M

LEVEL : Approx. 1 M

Insulation resistance

Min. 20 M

Ω at 500 V DC between the input terminal and ground.

Dielectric strength

Between input terminals

: 1000 V AC (50/60 Hz) for one minute

Between an input terminal and ground

: 1500 V AC (50/60 Hz) for one minute

Input source resistance

LEVEL

: 2 k

Ω or lower

Power consumption

Included in the main unit or the subunit to which the module is to be installed.

Advertising