Specifications

Measuring Accuracy and Resolution
Reference operation conditions

temperature: 23

±2°C; humidity: 55 ±10% RH; supply voltage 90 to 250 VAC (AC power supply)

/10 to 32 VDC (DC power supply); supply frequency: 50/60

±1% (AC power supply); warm-up

time: at least 30 minutes; when operating, the system must not adversely affect the operation of
other equipment.

Measured Data Item

Measuring Accuracy

Resolution

Effective voltage

± (0.5% of SPAN)

0.01 Vrms (for 25-V range), 0.1 Vrms (for 250-V range)

Effective current

± (0.5% of SPAN)

0.0001 Arms (for 0.5-A range), 0.001 Arms (for 5-A range)

Active power

± (1% of SPAN) φ = 0

0.01 W (for 12.5-W, 25-W and 37.5-W spans)

± (2.5% of SPAN) 0 <| φ |≤ 30

0.1 W (for 125-W, 250-W and 375-W spans)

± (5% of SPAN) 30 <| φ |≤ 80

1 W (for 1,250-W, 2,500-W and 3,750-W spans)

Apparent power

± (1% of SPAN) 0 ≤| φ |≤ 80

0.01 VA (for 12.5-VA, 25-VA and 37.5-VA spans)
0.1 VA (for 125-VA, 250-VA and 375-VA spans)
1 VA (for 1,250-VA, 2,500-VA and 3,750-VA spans)

Reactive power

± (5% of SPAN) 0 ≤| φ |< 60

0.01 Var (for 12.5-Var, 25-Var and 37.5-Var spans)

± (2.5% of SPAN) 60 ≤| φ |≤ 80

0.1 Var (for 125-Var, 250-Var and 375-Var spans)
1 Var (for 1,250-Var, 2,500-Var and 3,750-Var spans)

Frequency

±0.1Hz

0.01Hz

Power factor

± (2% of SPAN) 0 ≤| φ |≤ 80

0.01

Phase angle

±5deg 0 ≤| φ |≤ 80

0.1deg

Measuring Ranges

Measured Data Item

25 V-0.5A

25 V-5 A

250 V-0.5 A

250 V-5 A

Effective voltage

0.00 to 25.00 V rms

0.00 to 25.00 V rms

0.0 to 250.0 V rms

0.0 to 250.0 V rms

Vi (i = 1, 2, 3, 13, 0)

Effective current

0.0000 to 0.5000 A rms

0.000 to 5.000 A rms

0.0000 to 0.5000 A rms 0.000 to 5.000 A rms

Ii (i = 1, 2, 3, 13, 0)

Active power

–12.50 to 12.50 W

–125.0 to 125.0 W

–125.0 to 125.0 W

–1250 to 1250 W

P1, P2, P3

Active power

–25.00 to 25.00 W

–250.0 to 250.0 W

–250.0 to 250.0 W

–2500 to 2500 W

P13

Active power

–37.50 to 37.50 W

–375.0 to 375.0 W

–375.0 to 375.0 W

–3750 to 3750 W

P0

Apparent power

0.00 to 12.50 VA

0.0 to 125.0 VA

0.0 to 125.0 VA

0 to 1250 VA

VA1, VA2, VA3

Apparent power

0.00 to 25.00 VA

0.0 to 250.0 VA

0.0 to 250.0 VA

0 to 2500 VA

VA13

Apparent power

0.00 to 37.50 VA

0.0 to 375.0 VA

0.0 to 375.0 VA

0 to 3750 VA

VA0

Reactive power

0.00 to 12.50 Var

0.0 to 125.0 Var

0.0 to 125.0 Var

0 to 1250 Var

Var1, Var2, Var3

Reactive power

0.00 to 25.00 Var

0.0 to 250.0 Var

0.0 to 250.0 Var

0 to 2500 Var

Var13

Reactive power

0.00 to 37.50 Var

0.0 to 375.0 Var

0.0 to 375.0 Var

0 to 3750 Var

Var0

Power factor

–1.00 to 1.00

–1.00 to 1.00

–1.00 to 1.00

–1.00 to 1.00

PFi (i = 1, 2, 3, 13, 0)

Phase

–80.0 to 80.0 deg

–80.0 to 80.0 deg

–80.0 to 80.0 deg

–80.0 to 80.0 deg

PHi (i = 1, 2, 3, 13, 0)

Frequency

45.00 to 65.00 Hz

45.00 to 65.00 Hz

45.00 to 65.00 Hz

45.00 to 65.00 Hz

FREQ

4.4 Specifications of Power Monitor Module

Advertising