Installation and Wiring

DA100/DS400/DS600(when using DC power terminal connecter)

This applies only to products with power supply 2 suffix code.

Follow the warnings below to avoid electric shock or damaging the instrument.

WARNING

• Connect the power wires after checking that the power supply is turned

off to prevent electric shock.

• To prevent fire, use wires with cross sectional area of 0.3mm

2

(22AWG)

or more.

CAUTION

• If you connect the + and - terminals in reverse on the DA100/DS400/

DS600, the internal fuse will be blown (You cannot replace the fuse by
yourself. The instrument needs servicing in this case). If there is a
possibility of reversing the polarity, insert a fuse (rating shown below) in
the wiring. This will lower the chances of blowing the internal fuse.
Fuse: 250 V/T2.5 A to T4A (time lag), 20 mm glass tube fuse,

I

2

t = 12.5 to 32

(Recommended: A1350EF (250 V/T2.5 A, I

2

t=12.5: SCHURTER:

FST0034.3121))

Connecting procedure

1. Check thst the power switch is turned off.
2. Connect the accessory DC power terminal connector (part No. A1105JC) to the power wire

and the DA100/DS400/DS600. Use a power wire with a cross sectional area of 0.3 mm

2

(22AWG) or more.

21

Soldering

- (0 V DC)

+ (10 to 32 V DC)

DA100/DS400/DS600
Rated supply voltage: 12 to 28 V DC
Operating supply voltage: 10 to 32 V DC
Power consumption: About 25 VA max.

• DA100/DS400/DS600: DC power supply model

DC power terminal
connector

DC power
terminal

2.8 Connecting the Power Cord and Turning the Power ON/OFF