Yokogawa DA100 User Manual

Page 80

Advertising
background image

IM DA100-01E

4-11

4

Specifications

Insulation resistance

Min. 20M

Ω at 500V DC between the input terminal and ground.

Input bias current

max. 10nA

Dielectric strength

Between input terminals

: 1000V AC (50/60Hz) for one minute

Between an input terminal and ground

: 1500V AC (50/60Hz) for one minute

Input source resistance

DCV, TC

: 2k

Ω or lower

RTD

: 10

Ω or lower per line (Pt100 Ω)

5

Ω or lower per line (Pt50 Ω)

1

Ω or lower per line (Cu10 Ω)

Temperature coefficient

zero

: 0.01% of range/

°C

full span : 0.01% of range/

°C (0.02% of span/ °C for Cu10 Ω)

Thermocouple burnout

Detected in a thermocouple range (On/Off) enabled, current of 4

µA, detectable pulse width of

approx. 5ms
2 k

Ω or lower is considered “normal”

100 k

Ω or greater is considered “disconnected”

Detection interval for thermocouple burnout

2.5 sec. in a measurement period of 0.5 sec.
2 sec. in a measurement period of 2 sec.

Detection timing for thermocouple burnout

The burnout condition is detected at each measurement interval. However, with a 10ch universal
input module connected and a measurement interval of 0.5sec, the minimum interval of the
burnout detection becomes 2.5 sec.

Power consumption

Included in the main unit or the subunit to which the module is to be installed.

External dimensions and weight

Weight(kg)

0.5
0.5
1.0
1.0
1.5
1.5
0.5
0.5
1.0
1.0
1.5
1.5

External dimensions(W)x(H)x(D) mm

approx.57

x

137

x

88

approx.57

x

137

x

88

approx.114

x

137

x

88

approx.114

x

137

x

88

approx.171

x

137

x

88

approx.171

x

137

x

88

approx.57

x

137

x

88

approx.57

x

137

x

88

approx.114

x

137

x

88

approx.114

x

137

x

88

approx.171

x

137

x

88

approx.171

x

137

x

88

Type

DU100-11
DU100-12
DU100-21
DU100-22
DU100-31
DU100-32
DU200-11
DU200-12
DU200-21
DU200-22
DU200-31
DU200-32

4.2 Universal Input Module and DCV/TC/DI Input Module