Unable to run the ProCalV5 Data Source Edit Wizard

If the ProCalV5 Data Source Edit Wizard displays an error when you run it (by clicking
Start > All Programs > Rockwell Software > FactoryTalk AssetCentre Server >
ProCalV5 Data Source Configuration
), the cause may be that the FactoryTalk
AssetCentre Calibration Management server was not installed in the same directory as the
FactoryTalk AssetCentre server. To fix this problem, you must uninstall and then reinstall
the FactoryTalk AssetCentre server, making sure to install the FactoryTalk AssetCentre
Calibration Management server in the same directory as the FactoryTalk AssetCentre
server.