The following reevaluation possibilities are offered:
Set induction time manually
If this option is switched on then the
induction time
can be set
manually. This is done by pressing the [ Ctrl ] key and at the
same time moving the red vertical line in the
Live graph
window
with pressed-down left-hand mouse key to the required lo-
cation. The new manually set induction time is displayed in the
Manual
field and in the
Live graph
window.
Set stability time manually
If this option is switched on then the alteration in conductivity
Delta Kappa
for the calculation of the
stability time
can be al-
tered manually. After alterations in the
Delta Kappa
field the
newly calculated
Stability time
is shown both in the field with
the same name and in the
Live graph
window.
Set tangents manually
If this option is switched on then tangents can be set manually
to determine the
induction time
. The induction time is then de-
fined as being the intersection point of the tangents. The tan-
gents are set as follows:
Set first tangent