Metrohm 763 PVC Thermomat Manual User Manual
Page 94

4.7 Results
763 PVC Thermomat
87
Parameters
Parameters
tab of the
Determination and method data
window with the dis-
play of the measuring parameters used in the determination.
Temperature
Set temperature of sample.
Delta T
Temperature correction: Correction value for
increasing the block temperature in order to
reach the set temperature in the sample.
Gas flow
Gas flow through sample.
Start options
Start delay
Delay time until start of measurement.
Start mode
Mode for starting the channels:
per channel
All channels have been started individually.
per 2 channels
Each 2 channels have been started together.
per block
4 channels in one block have been started to-
gether.
Stop criteria
Time
Time until automatic termination of determina-
tion.
Conductivity
Conductivity limit. When this value is reached
the determination is automatically terminated.
Endpoint(s)
Automatic termination of the determination
when each active channel has reached all end-
points.