4 Operation

763 PVC Thermomat

70

Enter sample identifiers

The sample identifiers

ID1

and

ID2

can be entered in the corresponding

fields of the operating area in the control window before or after starting

the determination.

The entry for

ID1

can be selected from the automatically formed list of

identifiers which have already been entered. This list can be deleted

again with

763 PVC THERMOMAT CONTROL / Options / Clear sample ID his-

tory

.

Insert reaction vessels

When the heating block reaches the temperature defined in the method

(frame color of the

<Heat to...>

button turns green) the reaction vessels

can be inserted into the PVC Thermomat 763 as follows (see Fig. 4):

1 Insert reaction vessels

Place the prepared reaction vessels 11 in the heating block

openings on the 763 PVC Thermomat. Close any openings

which are not in use with stoppers or empty reaction vessels

to prevent contamination.

2 Connect reaction vessels

Connect the white connecting tubing 33 fastened to the reac-

tion vessel cover 22 to the corresponding connection piece 22

22

on the measuring vessel cover 55.

Screw the brown gas supply tubing 66 onto the reaction vessel

cover 22 and the corresponding connection 19

19 on the 763

PVC Thermomat (see Fig. 3).

Start determinations

Determinations can be started by clicking on the following buttons in

the operating area of the

763 PVC THERMOMAT CONTROL

window:

Starts a single determination for the corresponding
channel.
Starts a double determination for channels 1 and 2.

Starts a double determination for channels 3 and 4.

Starts a multiple determination for channels 1 to 4.

Starts all Block A channels.

Starts all Block B channels.