4.8 GLP
763 PVC Thermomat
125
Temperature
Conductivity
18 °C
127
µ
S/cm
19 °C
130
µ
S/cm
20 °C
133
µ
S/cm
21 °C
136
µ
S/cm
22 °C
138
µ
S/cm
23 °C
141
µ
S/cm
24 °C
144
µ
S/cm
25 °C
147
µ
S/cm
•
Click on
<Next>
in the GLP test window. The conductivity of
each channel is measured in sequence.
4 Confirm/cancel test results
•
When the last measurement has been completed the
TEST
RESULTS
(set values and measured values) for the measured
channels will be shown.
Test passed: The variation of the measured
conductivity is within the preset tolerance.
Test failed: The variation of the measured
conductivity is larger than the preset tolerance.
•
Check the results and click on
<Finish>
to save the GLP test
results or on
<Cancel>
to abort the GLP test without saving the
results.