Metrohm 763 PVC Thermomat Manual User Manual
Page 48

4.2 Instrument and software settings
763 PVC Thermomat
41
4.2.3
Timer
763 PVC THERMOMAT CONTROL / Tools / Timer function...
The
Automatic heater start
window opens in which the automatic start of
heating at a predetermined time can be entered independently for
block A or block B.
Start heater
Never
No automatic start of heating.
Once
Automatic start of heating at the time defined in
the
Start at...
field.
Daily
Automatic start of heating on the day selected
in the
<Which days>
field at the time defined in
the
Start at...
field.
<Which days>
This button opens the
Select days
window in
which the required day can be selected.
Start at...
If
Once
has been selected then the time and
date for the automatic start of heating can be
entered, if
Daily
has been selected the time.
4.2.4
Gas flow control
The flow of gas through the reaction vessels to the measuring vessels
can be switched on and off manually and shown in a separate window.
763 PVC THERMOMAT CONTROL / Tools / Gas flow control / Gas flow on
Switches on gas flow.
763 PVC THERMOMAT CONTROL / Tools / Gas flow control / Gas flow off
Switches off gas flow.