Metrohm 763 PVC Thermomat Manual User Manual
Page 133

4 Operation
763 PVC Thermomat
126
GLP gas flow test
763 PVC THERMOMAT CONTROL / Tools / GLP test / Gas flow
This menu item is used to start the automatic GLP test for gas flow
measurement. A gas flow meter is required for this test (not available
from Metrohm) so that the gas flow can be measured at each channel.
In the GLP test a separate window opens which guides you through the
test step-by-step. The next step is reached with
<Next>,
the previous
step with
<Back>
and the test is aborted with
<Cancel>
.
Proceed as follows to carry out the GLP gas flow test:
1 Select instrument
•
Select unit 1...4 in the control window for which the GLP gas
flow test has to be carried out.
2 Start GLP test
•
Click on
763 PVC THERMOMAT CONTROL / Tools / GLP test / Gas
flow
and click on
<Next>
in the GLP test window.
Gas flow rate
Gas flow in L/h.
•
Enter the
Gas flow rate
and click on
<Next>
.