Metrohm 763 PVC Thermomat Manual User Manual
Page 124
4.8 GLP
763 PVC Thermomat
117
4.8.2
GLP properties
763 PVC THERMOMAT CONTROL / Tools / GLP test / Properties
This menu item is used to open the
GLP properties
window for the se-
lected instrument; this contains the tabs for temperature, conductivity
and gas flow.
Enable GLP monitoring
Activates automatic
GLP monitoring.
GLP alert interval
Number of days before next GLP test is due.
Tolerance
States the tolerance values for the particular
GLP test. If the result is within the
Set value
±±
Tolerance
range then the test has been passed.
By clicking on
<Set default>
the preset default
value is entered in this field.
Additional alert message text
Additional text which is to be added to the re-
quest to carry out the next GLP test.
4.8.3
GLP status
763 PVC THERMOMAT CONTROL / Tools / GLP test / Status
Opens the
GLP status
window for the selected instrument.