4 Operation
763 PVC Thermomat
56
The
Method parameters
window contains the following buttons:
<Save>
Saves method under the same name.
<Save as>
Saves method under a new name; the old
method is retained.
<Print>
Prints out method report.
<Cancel>
Closes method parameter window without
saving the modified parameters.
Parameters
Tab
Parameters
of the
Method parameters
window with measuring pa-
rameters.
Temperature
Set temperature of the sample.
Range:
50...220 °C
Delta T
Temperature correction: Correction value for
increasing the block temperature in order to
reach the set temperature in the sample. This
value is determined by measuring the variation
of the actual sample temperature from the
block temperature using a calibrated external
temperature sensor (see section 4.4.2).
Range:
-9.9...9.9 °C