4 Operation
763 PVC Thermomat
48
4.4
Calibration functions
4.4.1
Determination of cell constants
As in PVC Thermomat measurements only the alterations in the con-
ductivity are measured and evaluated, the cell constants of the con-
ductivity cells used in the PVC Thermomat, which are normally in the
range
1.00...1.20
, only need to be entered exactly if the absolute value
for the conductivity is to be shown correctly. The exact entry is also re-
quired before carrying out the GLP test for conductivity. The cell con-
stants can either be entered manually or determined automatically with
the aid of a defined standard solution.
Manual entry of cell constants
763 PVC THERMOMAT CONTROL / Tools / Cell constants / Manual...
The
Enter/check cell constants
window appears in which the cell constant
for the conductivity cell can be entered for each
Channel
.
Automatic determination of cell constants
763 PVC THERMOMAT CONTROL / Tools / Cell constants / Automatic...
The
Determine cell constants
window appears: