4 Operation

763 PVC Thermomat

48

4.4

Calibration functions

4.4.1

Determination of cell constants

As in PVC Thermomat measurements only the alterations in the con-

ductivity are measured and evaluated, the cell constants of the con-

ductivity cells used in the PVC Thermomat, which are normally in the

range

1.00...1.20

, only need to be entered exactly if the absolute value

for the conductivity is to be shown correctly. The exact entry is also re-

quired before carrying out the GLP test for conductivity. The cell con-

stants can either be entered manually or determined automatically with

the aid of a defined standard solution.

Manual entry of cell constants

763 PVC THERMOMAT CONTROL / Tools / Cell constants / Manual...

The

Enter/check cell constants

window appears in which the cell constant

for the conductivity cell can be entered for each

Channel

.

Automatic determination of cell constants

763 PVC THERMOMAT CONTROL / Tools / Cell constants / Automatic...

The

Determine cell constants

window appears: