Metrohm 763 PVC Thermomat Manual User Manual
Page 99

4 Operation
763 PVC Thermomat
92
Automatic report output
Send to printer immediately
Automatic printout of the reports defined under
Include in report
following the end of the deter-
mination.
Send to file
Automatic output of determination and method
data following the end of the determination as a
text file
*.txt
.
Directory
Directory in which the report will be stored.
Warnings
Warnings
tab of the
Determination and method data
window with the dis-
play of warnings about the determination.
Description
If the determination was not carried out properly
then the following warnings are possible:
Temperature not reached
Set temperature was not reached at start of
determination.
Missing measuring points (start time)
Measuring points are missing from the given
time (in s) onwards.
Missing measuring points (end time)
Measuring points are again present from the
given time (in s) onwards.
Manual stop
The determination was stopped manually.
Abortion after communication error
Automatic determination stop because con-
nection between the 763 PVC Thermomat and
PC was interrupted.
Value
Value of the warning (e.g. temperature, time).