4 Operation
763 PVC Thermomat
122
Ext. temperature
The temperature measured by the external
temperature sensor.
Measuring time
The elapsed time since the start of the tem-
perature measurement.
•
When the measurement has been carried out you are re-
quested to insert the external temperature sensor together
with the GLP insert in the next selected channel and to start
the measurement. The change must be carried out quickly.
•
Carry out the GLP temperature measurement for each se-
lected channel in sequence.
5 Confirm/cancel test results
•
When the last temperature measurement has been com-
pleted the
TEST RESULTS
for the measured channels are
shown.
Test passed: The variation of the measured
temperature is within the preset tolerance.
Test failed: The variation of the measured
temperature is larger than the preset tolerance.
•
Check the results and click on
<Finish>
to save the GLP test
results or on
<Cancel>
to abort the GLP test without saving the
results.