6 Appendix
763 PVC Thermomat
136
6.1.3
Conductivity measurement
Sensor
6.0913.130 Conductivity cell built into
measuring vessel cover
Construction
Conductivity measurement cell with 2 steel
electrodes
Measurement principle
Alternating current measurement with 1 kHz
frequency and approx. 1.7 V amplitude
(peak to peak).
Cell constant
1.00
…
1.20
(the exact value can be manually entered or
automatically determined)
Measurement range
0
…
400
µ
S/cm
Resolution
0.1
µ
S/cm
Maximum error
±
1 %
±
0.5
µ
S/cm
6.1.4
Gas flow regulation
Gas flow supply
External supply from N
2
cylinder
Adjustable range
7
…
25 L/h
Max. variation from
set range
±
10 %
6.1.5
GLP test set
External temperature sensor
Type
Pt100 jacketed element 1.9 mm dia.,
with 4-lead technology; DIN IEC 751 class B
Calibration accuracy
0
…
90 °C:
±
0.02 °C
90
…
220 °C:
±
0.03 °C
Max. variation of measured
temperature from true value
(50
…
220 °C)
±
0.3 °C
Test resistor
Resistance
10 k
Ω
6.1.6
RS232 interface
Connector
Dsub 9-pin (male)
Function
TxD and RxD signal for connection with
software handshake
Default settings
9600 baud, 8 bit, 1 stop bit, no parity,
XON/XOFF