Metrohm 763 PVC Thermomat Manual User Manual

Page 135

Advertising
background image

4 Operation

763 PVC Thermomat

128

4.8.5

GLP results

Print GLP certificate

763 PVC THERMOMAT CONTROL / Tools / GLP test / Print GLP results

This menu item is used to print out the GLP certificate for the selected

instrument; this contains the printout date, the

Serial number

of the in-

strument and the results of the previous GLP tests carried out on this

instrument. Each of the GLP tests for temperature, conductivity and gas

flow contains the following information:

Date

Date of the last GLP test carried out.

Tester

Operator who carried out the GLP test.

Test result

Information about whether the test was passed

or failed.

In order to print out the test results

GLP monitoring

must be enabled

(see section 4.8.2).

View GLP results

763 PVC THERMOMAT RESULTS / View / GLP

The

GLP

selection window opens to provide three possible selections:

<Temperature>

Opens the overview of the results contained in

the current database of the GLP test for tem-

perature measurement.

<Conductivity>

Opens the overview of the results contained in

the current database of the GLP test for con-

ductivity measurement.

Advertising