4.4 Calibration functions
763 PVC Thermomat
49
Block A
Channel 1...4
Each channel
of Block A for which the cell con-
stant of the conductivity cell is to be determined
can be marked.
Block B
Channel 1...4
Each channel
of Block B for which the cell con-
stant of the conductivity cell is to be determined
can be marked.
Conductivity of standard solution
Conductivity of the standard solution in
µ
S/cm
which is filled into all measuring vessels for
which the cell constant of the conductivity cell
is to be determined.
A solution of c(KCL) = 1 mmol/L, prepared
from the 6.2301.060 Conductivity standard (KCl
0.1 mol/L) available as an option, can be used
as standard solution. The conductivity of this
standard solution is:
Temperature
Conductivity
18 °C
127
µ
S/cm
19 °C
130
µ
S/cm
20 °C
133
µ
S/cm
21 °C
136
µ
S/cm
22 °C
138
µ
S/cm
23 °C
141
µ
S/cm
24 °C
144
µ
S/cm
25 °C
147
µ
S/cm
<Start>
Starts the automatic determination of the cell
constants. When the determination has ended
the
Enter/check cell constants
window appears in
which the automatically determined cell con-
stants have been entered:
Click on
<OK>
to accept these values or on
<Cancel>
to retain the previous values.