3 cal meas conc - measuring parameters – Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 1038

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1022

■■■■■■■■

tiamo 2.1 (ProcessLab)

Selection

1 | 2 | 3 | 4 | off

Default value

1

off
Means that no stirrer will be used.

Stirring rate

Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.

Range

-15 ... 15

Default value

8

Switch off automatically

on | off (Default value: on)

If this check box is activated, the stirrer will be switched off automatically
when the command has finished.

5.6.5.7.3

CAL MEAS Conc - Measuring parameters

Tab: Method

CAL MEAS Conc Properties... Measuring parameters

Command name

Name of the command.

Input

25 characters

The measuring parameters for the calibration are defined on this tab.

Measurement with drift control
The ISE calibration is carried out drift-controlled. The measurement is stop-
ped as soon as the Signal drift is reached or a defined waiting time has
elapsed.

Measured value drift

The measured value is only accepted if the signal drift defined here has been
fallen short of.

Range

0.1 ... 999.0 mV/min

Default value

2.0 mV/min

Min. waiting time

The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The drift continues
to be checked while the waiting time is elapsing.

Range

0 ... 999999 s

Default value

10 s