5 Method

tiamo 2.1 (ProcessLab)

■■■■■■■■

733

not defined
No tests will be carried out.

Sensor

Measuring input

Selection of the measuring input to which the sensor is connected.

Titrando
Selection

1 | 2

Default value

1

Sensor

Selection of a sensor of type Metal electrode from the sensors configured
in the sensor table. The calibration data for the sensor will be adopted for
the determination.

Selection

Sensor name | Metal electrode

Default value

Metal electrode

I(pol)

The polarization voltage is the voltage applied to the polarizable electrode
during a amperometric measurement.

Titrando, 855
Range

-125.0 ... 125.0 µA (Increment: 0.5)

Default value

5.0 µA

Titrino
Range

-127 ... 127 µA (Increment: 1)

Default value

5 µA

Electrode test

on | off (Default value: off)

If this option is enabled, then an electrode test will be carried out for polar-
izable electrodes during the transition from an inactive condition to a
measurement. A check is also made while doing so to ensure that the elec-
trode is properly connected and that no short-circuit is present.

Temperature measurement

Type of temperature measurement.

Titrando, 855
Selection

continuous | automatic | off

Default value

automatic