5 Method

tiamo 2.1 (ProcessLab)

■■■■■■■■

1059

Stop determination
The running DOS command is quit, then the exit track (if present) is
started and the determination is stopped.
Stop command (only Titrando)
The running DOS command is quit, then the next command is executed.
Wait for [Continue]
Reagent dosing in the current DOS command is interrupted and a mes-
sage appears. As soon as the monitored measured value is again within
the limits (including hysteresis) reagent dosing can be resumed by press-
ing the [Continue] button in this message box.
Wait for limit ok
Reagent dosing in the current DOS command is interrupted. As soon as
the monitored measured value is again within the limits (including hys-
teresis), reagent dosing will be resumed automatically.
none
No action will be taken if limits are exceeded.

Monitoring Temperature

Monitoring Temperature

on | off (Default value: off)

If this option is switched on then the temperature will be monitored and any
exceeded limits will be entered in the measuring point list.

Lower limit

Lower limit of the temperature. When the temperature falls below this value,
the action Lower temperature limit exceeded is triggered.

Titrando, 855
Range

-20.0 ... 150.0 °C

Default value

-20.0 °C

Titrino
Range

-170.0 ... 500.0 °C

Default value

-170.0 °C

Lower hysteresis

Lower hysteresis of the temperature. When the temperature once again
exceeds the lower limit by this hysteresis value, the action Temperature
lower limit ok
is triggered.

Titrando, 855
Range

0.0 ... 150.0 °C

Default value

0.2 °C