5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

814

■■■■■■■■

tiamo 2.1 (ProcessLab)

Selection

maximum

Default value

maximum

Titrino
Range

0.01 ... 150.00 mL/min

Selection

maximum

Default value

maximum

Pause 2

Pause 2

Waiting time, e.g. for the electrode to settle down after the start or a reac-
tion time after the addition of the start volume. The pause follows at the
end of all the start conditions.

Range

0 ... 999999 s

Default value

0 s

5.6.3.8.6.4

STAT pH - Control parameters

Tab: Method

STAT pH Properties... Control parameters

Command name

Name of the command.

Input

25 characters

Control point

Control point at pH

Definition of the pH value to be controlled.

Titrando, 855
Range

-20,000 ... 20,000

Selection

off

Default value

off

Titrino
Range

-20.00 ... 20.00

Selection

off

Default value

off

Titration rate

Titration rate

Three predefined rates, for which the parameters for Control are not dis-
played, can selected for the titration rate. The User setting must be selected
in order to be able to edit these parameters.