5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1180
■■■■■■■■
tiamo 2.1 (ProcessLab)
Distance
Distance to be moved to.
Visible only in cases of stepping motors where Type = Linear drive.
Range
-214783647 ... 214783647 mm
Default value
0 mm
Angle
The angle to be moved to.
Not visible in cases of stepping motors where Type = Rotary drive.
Range
-214783647 ... 214783647 °
Default value
0 °
Note
The moving direction of the stepping motor depends on the algebraic
sign of the value for Distance or Angle:
■
≥0: Move in positive direction.
■
<0: Move in negative direction.
Start delay
Waiting time in seconds or minutes before the initiated action is started.
Range
0 ... 999999
Default value
0
Selection
s | min
Default value
s
Timeout
Time in seconds or minutes within which the action must have run through
completely. If this is not done after the maximum waiting time has expired,
the status of the stepping motor will be set to ERROR.
Range
0 ... 999999
Default value
10
Selection
s | min
Default value
s
End delay
Waiting time, after expiry of maximum waiting time
Range
0 ... 999999
Default value
0