Metrohm tiamo 2.1 (ProcessLab) User Manual
Page 1512

8.4 Manual Control - Functions
■■■■■■■■■■■■■■■■■■■■■■
1496
■■■■■■■■
tiamo 2.1 (ProcessLab)
Start delay
Waiting time in seconds or minutes before the initiated action is started.
Range
0 ... 999999
Default value
0
Selection
s | min
Default value
s
Working time
Time in seconds or minutes for which the digital output is set to ACTIVE.
After the working time has elapsed, it is set to INACTIVE again.
Range
0 ... 999999
Default value
0.01
Selection
s | min
Default value
s
End delay
Waiting time after the working time has expired
Range
0 ... 999999
Default value
0
Selection
s | min
Default value
s
Status
Display of the current status.
Selection
ACTIVE | INACTIVE | ERROR
Time to finish
Display of the time to finish of the timer control.
[Start]
Sets the digital output to ACTIVE.
[Stop]
Sets the digital output to INACTIVE.
Set default
Port description
Display of the port description.
Port
Display of the internal description of the port.