8.4 Manual Control - Functions
■■■■■■■■■■■■■■■■■■■■■■
1484
■■■■■■■■
tiamo 2.1 (ProcessLab)
Selection
On/Off | Duration
Default value
On/Off
Duration
Enter the duration during which the pump/valve should remain switched on.
This field only appears for Operation = Duration.
Range
1 ... 999999 s
Default value
60 s
[Start]
Start pump 2. In the status display the running time since start is shown.
[Stop]
Stop pump 2-
8.4.4.6
Sample changer - Heater/Gas
Program part: Manual Control
Here the heater and gas flow on the 774 and 874 Oven Sample Processor
can be switched on and off.
Heater
Heat the oven up to the desired temperature during the defined heating up
time.
Temperature
Temperature up to which heating should take place. With Init it is heated
up to the initial temperature set on the device.
Range
50 ... 250 °C
Default value
50 °C
Selection
Init
Heating up time
Duration of the heating up period until reaching the desired temperature.
Range
1 ... 999 min
Selection
off
Default value
off
[Start]
Start heater. In the status display the current temperature is shown.
[Stop]
Stop heater.