2 stat - control range – Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 817

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.1 (ProcessLab)

■■■■■■■■

801

Commands
Depending on the measured value, one of the following twoSTAT com-
mands can be chosen:

STAT pH
STAT titration with pH electrodes (measured quantity pH)

STAT U
STAT Titration with metal electrodes (measured quantity U).

5.6.3.8.2

STAT - Control range

Tab: Method

STAT Properties... Control parameters

Set a large control range for steep curves and a small control range for flat
ones. A good approximation for the start of the control range is given by
the point where the tangents intersect.

Advertising