5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1034

■■■■■■■■

tiamo 2.1 (ProcessLab)

5.6.5.9.3

CAL Spec - Measuring parameters

Tab: Method

CAL Spec Properties... Measuring parameters

Command name

Name of the command.

Input

25 characters

The parameters for the measuring procedure can be defined on this tab.

Integration time

Integration time for spectrum recording.

Range

0.01 ... 600000 ms

Default value

6 ms

Averaged spectra

The number of spectra that are recorded and averaged.

Range

1 ... 10000

Default value

10

5.6.5.9.4

CAL Spec - Calibration wavelengths

Tab: Method

CAL Spec Properties... Calibration wavelengths

Command name

Name of the command.

Input

25 characters

The reference wavelengths with which the spectrometer is calibrated can
be entered on this tab.

Tolerance

Tolerance for the preset wavelength when assigning the highest peak max-
imum found during the calbration measurement to the reference wave-
length. If a peak is outside the range Wavelength +/- Tolerance, the
calibration will be canceled. The existing calibration in the configuration
remains in effect.

Range

0.1 ... 1100.0 nm

Default value

0.5 nm

[New]

Opens the dialog window Calibration wavelength # to enter wave-
lengths and tolerances (see Chapter 5.6.5.9.5, page 1035).