8 kft ipol - additional evaluations, Additional evaluations – Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 761

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.1 (ProcessLab)

■■■■■■■■

745

Note

Conditioning must already be completed (= Condok) at the time the
SEND command is received, otherwise the command to start the titra-
tion will be ignored.

Show measured value during conditioning

on | off (Default value: off)

If this option is enabled, then the measured value will be shown in the sub-
window Live display of the program part Workplace, in addition to the
drift and the time.

5.6.3.5.2.8

KFT Ipol - Additional evaluations

Tab: Method

▶ KFT Ipol ▶ Properties... ▶ Additional evaluations

Command name

Name of the command.

Input

25 characters

The following additional methods for the evaluation of the titration curves
can be activated and defined on these tabs:

Fix endpoint evaluation

Fix endpoint evaluation

on | off (Default value: off)

If this check box is activated, then the associated values for the other meas-
ured quantities will be interpolated with a fixed measured quantity (Meas-
ured value
, Volume or Time) for the fix endpoint from the measuring
point list.

Moves the selected line up (changes the sequence).

Moves the selected line down (changes the sequence).

[New]

Opens the dialog window Fix endpoint evaluation # to enter a new fixed
endpoint (see Chapter 5.6.3.5.2.10, page 748).

[Properties]

Opens the dialog window Fix endpoint evaluation # to edit the fixed
endpoint selected in the table (see Chapter 5.6.3.5.2.10, page 748).

Advertising