5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

848

■■■■■■■■

tiamo 2.1 (ProcessLab)

Titrino
Range

0.00 ... 150.00 mL/min

Default value

0.00 mL/min

Lower hysteresis

The lower hysteresis represents a tolerance range for the lower limit of the
dosing rate. When the dosing rate exceeds the lower limit by this hysteresis
value, the action Dosing rate lower limit ok is triggered.

Titrando, 855
Range

0.00 ... 166.00 mL/min

Default value

0.20 mL/min

Upper limit

Upper limit of the dosing rate. If the dosing rate exceeds this limit, then the
event Dosing rate upper limit exceeded will be triggered.

Titrando, 855
Range

0.00 ... 166.00 mL/min

Default value

166.00 mL/min

Titrino
Range

0.00 ... 150.00 mL/min

Default value

150.00 mL/min

Upper hysteresis

Upper hysteresis of the dosing rate. When the dosing rate again falls below
the upper limit by this hysteresis value, then the event Dosing rate upper
limit ok
is triggered.

Titrando, 855
Range

0.00 ... 166.00 mL/min

Default value

0.20 mL/min

Action

Selection of the event to be triggered if the upper or lower limit is exceeded:

Selection

Stop determination | Stop command (Titrando,
855)
| Wait for [Continue] | Wait for limit ok |
none

Default value

none

Stop determination
The running STAT command is quit, then the exit track (if present) is
started and the determination is stopped.