2 minimum/maximum evaluation, 3 break-point evaluation – Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 872

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

856

■■■■■■■■

tiamo 2.1 (ProcessLab)

5.6.3.9.2

Minimum/Maximum evaluation

Parameters: Method

▶ DET/MET/SET/KFT/KFC/MEAS ▶ Properties... ▶ Additional

evaluations

▶ Minimum/Maximum evaluation

For the minimum or maximum measured value the associated volume,
time and temperature are interpolated from the measuring point list. The
evaluation begins as soon as the slope of the curve exceeds a particular
threshold value.

5.6.3.9.3

Break-point evaluation

Parameters: Method

▶ DET/MET/MEAS ▶ Properties... ▶ Additional evalua-

tions

▶ Break-point evaluation

A break-point evaluation is used to determine sharp change of direction in
the titration curve. This evaluation is primarily used for photometric and
conductivity titrations. The method is based on the search for extremes in
the second derivative of the curve.

Advertising