3 measure - intensity – Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 1505

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

8 Manual Control

tiamo 2.1 (ProcessLab)

■■■■■■■■

1489

Sample solution

Select name from the table Sample solution (TC conductivity).

Selection

DIN | Entries of the table

Default value

DIN

[Start]

Start measuring.

[Stop]

Stop measuring.

8.4.5.3

Measure - Intensity

Program part: Manual Control

Here the intensity of an uncorrected spectrum can be measured and dis-
played.

Start wavelength

Lower limit of the spectrum.

Range

200.0 ... 1100.0 nm

Default value

400.0 nm

End wavelength

Upper limit of the spectrum.

Range

200.0 ... 1100.0 nm

Default value

1000.0 nm

Integration time

Integration time for spectrum recording.

Range

0.01 ... 600000 ms

Default value

6 ms

Averaged spectra

The number of spectra that are recorded and averaged.

Range

1 ... 10000

Default value

10

Smoothing

Number of neighboring pixels with which the value for each measurement
pixel is determined.

Range

0 ... 100 Pixels

Default value

0 Pixels