6 met u - potentiometric evaluation – Metrohm tiamo 2.1 (ProcessLab) User Manual

Page 594

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

578

■■■■■■■■

tiamo 2.1 (ProcessLab)

Titrando, 855
Range

0 ... 999999 s

Selection

off

Default value

off

Filling rate

Speed with which the dosing cylinder is to be refilled after the titration. The
maximum filling rate depends on the cylinder volume of the exchange unit
or dosing device used.

Titrando, 855
Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum

Titrino
Range

0.01 ... 150.00 mL/min

Selection

maximum

Default value

maximum

5.6.3.3.3.6

MET U - Potentiometric evaluation

Tab: Method

MET U Properties... Potentiometric evaluation

Command name

Name of the command.

Input

25 characters

Parameters for the potentiometric evaluation of the titration curve with
automatic equivalence point recognition are defined on this tab. One of the
following three evaluation methods can be chosen here:

Potentiometric evaluation

Selection

Evaluation without window | Evaluation with
measured value window (U)
| Evaluation with
volume window (mL)

Default value

Evaluation without window

Evaluation without window
With this option set, the two parameters EP criterion and EP recog-
nition
will be applied across the entire range of the titration curve.

Advertising