Procedure – LaMotte SMART3 Colorimeter User Manual

Page 130

Advertising
background image

PROCEDURE

1. Press and hold

until colorimeter turns on.

2. Press

ENTER

to select TESTING MENU.

3. Select

ALL TESTS (or another sequence containing 019 Chlorine Diox) from

TESTING MENU.

4. Scroll to and select 019 Chlorine Diox from menu.
5. Rinse a clean tube (0290) with sample water. Fill to 10 mL line with sample.
6. Insert tube into chamber, close lid and select SCAN BLANK.
7. Remove tube from colorimeter. Add 5 drops of Glycine Solution (6811).
8. Add one *Chlorine DPD #1 Instrument Grade Tablet (6903A). Cap and shake

for 10 seconds. Invert 5 times slowly. Solution will turn pink if chlorine dioxide is

present.

9. Immediately insert tube into chamber, close lid and select SCAN SAMPLE.
10. Press

to turn colorimeter off or press

EXIT

to exit to a previous menu or

make another menu selection.

CHLORINE DIOXIDE

SMART3 Test Procedures 11.10

Test P

rocedures

Advertising