PROCEDURE

1. Press and hold

until colorimeter turns on.

2. Press

ENTER

to select TESTING MENU.

3. Select

ALL TESTS (or another sequence containing 028 Cu BCA Tablet) from

TESTING MENU.

4. Scroll to and select 028 Cu BCA Tablet from menu.
5. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with sample.
6. Insert tube into chamber, close lid and select SCAN BLANK.
7. Remove tube from colorimeter and add one *Copper Tablet (T-3808). Cap and

shake vigorously until tablet dissolves. Solution will turn purple if copper is

present. Wait 2 minutes.

8. At end of 2 minute waiting period, mix, insert tube into chamber, close lid and

select SCAN SAMPLE. Record result.

9. Press

to turn colorimeter off or press

EXIT

to exit to a previous menu or

make another menu selection.

COPPER, BCA, Tablet

SMART3 Test Procedures 11.10