Procedure – LaMotte SMART3 Colorimeter User Manual
Page 154
PROCEDURE
1. Press and hold
until colorimeter turns on.
2. Press
ENTER
to select TESTING MENU.
3. Select
ALL TESTS (or another sequence containing 028 Cu BCA Tablet) from
TESTING MENU.
4. Scroll to and select 028 Cu BCA Tablet from menu.
5. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with sample.
6. Insert tube into chamber, close lid and select SCAN BLANK.
7. Remove tube from colorimeter and add one *Copper Tablet (T-3808). Cap and
shake vigorously until tablet dissolves. Solution will turn purple if copper is
present. Wait 2 minutes.
8. At end of 2 minute waiting period, mix, insert tube into chamber, close lid and
select SCAN SAMPLE. Record result.
9. Press
to turn colorimeter off or press
EXIT
to exit to a previous menu or
make another menu selection.
COPPER, BCA, Tablet
SMART3 Test Procedures 11.10
Test P
rocedures