other Total Nitrogen Acid Reagent Tube. This is the sample.

22. Cap the tubes and invert 10 times to mix.

CAUTION: The tubes will be hot.

Note: Invert slowly and completely for accurate results. Hold tubes with caps up.

Invert the tube and wait for the air bubble to fl ow to the bottom of the tube. Turn

the tube upright and wait for the air bubble to return to the top of the tube. This

is one inversion.

23. Wait 5 minutes.
24. Wipe the tubes with a damp towel to remove fi ngerprints and smudges. Wipe

with a dry towel.

25. Insert the blank tube into the chamber. Select SCAN BLANK. Remove the blank

tube from the colorimeter.

26. Insert the sample tube into the chamber. Select SCAN SAMPLE. Record the

result as Total Nitrogen in mg/L N.

27. Press

to turn the colorimeter off or press

EXIT

to exit to a previous menu

or make another menu selection.

NOTES: For greater accuracy, use laboratory grade pipets.

To order reagent refi lls, Order Code R-4026.

NITROGEN, Total