Procedure–low range – LaMotte SMART3 Colorimeter User Manual

Page 242

Advertising
background image

PROCEDURE–LOW RANGE

A. PREPARATION OF HR REAGENT

NOTE: The quantity of Indigo Blue Stock solution (3989) supplied will prepare one

batch of HR Reagent for the High Range Ozone procedure or fi ve batches of HR

Reagent for the Low Range Ozone procedure.

1. Use the 50 mL graduated cylinder to carefully add 45 mL of *Ozone Buffer

(3991) to amber glass bottle marked HR Reagent (3989-MT).

2. Use the 5 mL transfer pipet (0329) and pump (30527) to add 5 mL of Indigo

Blue Stock Solution (3989) to the amber glass bottle. Cap and mix.

B. DETERMINATION OF OZONE
3. Use the 1.0 mL transfer pipet (2-2170) and pump (30527) to add 1.0 mL of HR

Reagent (3989) to each of 2 clean tubes (0290).

4. If chlorine is present add 3 drops Chlorine Inhibitor (3990) to each tube. Cap

tubes.

5. Take one of the prepared tubes (0290) and sampling apparatus (0681) to

sampling site.

6. Lower end of tubing of sampling apparatus to desired depth. Slowly withdraw

and depress plunger several times to purge syringe and tubing. Slowly

withdraw plunger to fi ll purged syringe.

7. Remove plastic tubing from syringe. Remove cap from the prepared tube.

Place tip of syringe against inside of the prepared tube. Slowly depress plunger

and fi ll to the 10 mL line and cap. This is the Sample Tube.

NOTE: DO NOT SHAKE OR INVERT THE SAMPLE.

8. Fill the second prepared tube (0290) to the 10 mL line with ozone free water.

This is the Reagent Blank.

9. Press and hold

until colorimeter turns on.

10. Press

ENTER

to select TESTING MENU.

11. Select ALL TESTS (or another sequence containing 071 Ozone LR) from

TESTING MENU.

12. Scroll to and select 071 Ozone LR from menu.
13. Insert the Reagent Blank tube into chamber, close lid and select SCAN

BLANK.

14. Insert reacted Sample Tube into chamber, close lid and select SCAN SAMPLE.

Record result.

15. Press

to turn colorimeter off or press

EXIT

to exit to a previous menu or

make another menu selection.

NOTE: HR Reagent must be made fresh each week. If reagent is refrigerated, it

may be kept up to 3 weeks.

OZONE, Indigo

SMART3 Test Procedures 11.11

Test P

rocedures

Advertising