APPLICATION:

Boilers

RANGE:

0.000–0.700 ppm DEHA (Diethylhydroxylamine)

0.000–0.900 ppm Carbohydrazide

0.00–3.00 ppm Erythorbic Acid

0.00–2.00 ppm Hydroquinone

0.00–3.00 ppm Methylethylketoxime

MDL:

0.004 ppm DEHA

0.01 ppm Carbohydrazide

0.02 ppm Erythorbic Acid

0.01 ppm Hydroquinone

0.01 ppm Methylethylketoxime

METHOD:

Ferric iron is reduced to ferrous iron by oxygen

scavengers in proportion to the concentration in the

sample. The resulting ferrous iron reacts with an indicator

to produce a purple color.

SAMPLE HANDLING

& PRESERVATION:

Analyze samples immediately. Rinse sample containers

and glassware with 1:1 hydrochloric acid to avoid iron

contamination.

INTERFERENCES:

Other oxygen scavengers, such as DEHA,

carbohydrazide, erythorbic acid, hydroquinone and

methylethylketoxime will interfere. Stray light and

substances which complex iron or reduce ferric iron will

also interfere.