Procedure – LaMotte SMART3 Colorimeter User Manual

Page 158

Advertising
background image

PROCEDURE

1. Press and hold

until colorimeter turns on.

2. Press

ENTER

to select TESTING MENU.

3. Select

ALL TESTS (or another sequence containing 031 Cu Thiocarbamate)

from TESTING MENU.

4. Scroll to and select 031 Cu Thiocarbamate from menu.
5. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with sample.
6. Insert tube into chamber, close lid and select SCAN BLANK.
7. Remove tube from colorimeter and add 5 drops of *Copper 1 (6446). Cap and

mix. Solution will turn yellow if copper is present.

8. Insert tube into chamber, close lid and select SCAN SAMPLE. Record result.
9. Press

to turn colorimeter off or press

EXIT

to exit to a previous menu or

make another menu selection.

NOTE: The reaction may stain the tubes. Scrub the tubes thoroughly after each

use.

COPPER, Thiocarbamate

SMART3 Test Procedures 11.10

Test P

rocedures

Advertising