Procedure – LaMotte SMART3 Colorimeter User Manual

Page 206

Advertising
background image

PROCEDURE

1. Press and hold

until colorimeter turns on.

2. Press

ENTER

to select TESTING MENU.

3. Select

ALL TESTS (or another sequence containing 056 Manganese HR)

from TESTING MENU.

4. Scroll to and select 056 Manganese HR from menu.
5. Rinse a tube (0290) with sample water. Fill to the 10 mL line with sample.
6. Insert tube into chamber, close lid and select SCAN BLANK.
7. Remove tube from colorimeter. Use the 0.1 g spoon (0699) to add two

measures of Manganese Buffer Reagent (6310). Cap and mix until powder

dissolves.

8. Use the 0.15 g spoon (0727) to add one measure of *Manganese Periodate

Reagent (6311). Cap and shake for one minute. An undissolved portion of the

reagent may remain in the bottom of the tube without adversely affecting the

test results. Wait two minutes for maximum color development. Solution will

turn pink if manganese is present.

9. At the end of the two minute waiting period, mix, insert tube into chamber,

close lid and select SCAN SAMPLE. Record result.

10. Press

to turn colorimeter off or press

EXIT

to exit to a previous menu or

make another menu selection.

MANGANESE, High Range

SMART3 Test Procedures 11.10